Part of the Oxford Instruments Group
Expand

Technical Articles

Camera Vacuum Flanges for High Energy Detection | An Overview

Author: Andor

Published: 10 Feb 2021 · Last updated: 11 Sep 2025

In order to efficiently detect photons with energy in the vacuum and extreme ultraviolet and soft and hard x-ray range, an environment free of particles that could interfere with the useful photon signal is generally created between the high energy radiation source and the detector. This is achieved by pumping out of the experimental chamber the atomic or molecular species that would absorb the useful radiation and be ionised, hence creating ‘pollutant’ species.

Here we highlight the key vacuum flange interfaces available for connecting Andor CCD detectors to vacuum chambers or spectrographs for a range of applications including x-ray plasma diagnostics, extreme ultraviolet (EUV/XUV) imaging & microscopy, x-ray diffraction imaging or spectroscopy, or high harmonic generation (HHG).

Energy range considerations

A high energy photon can be absorbed by molecules or atoms present in the experiment environment, and ionise these species at the higher energies. The regions relevant for the present discussion are defined as follows [1]:

Radiation domain Wavelength range (nm) Photon energy range (keV) Comments
Ultraviolet
Vacuum UV (VUV)
Extreme UV (EUV)
 
10 – 200
10 - 121
 
0.124 – 0.006 0.124 – 0.1
Absorbed by atmospheric oxygen, increasingly pronounced <~180 nm, potential ionising radiation
Soft X-Ray
Extreme UV (XUV)
 
0.1 – 10
 
12.4 – 0.124
 
Ionising radiation
Hard X-Ray 0.001 – 0.1 1,240 - 12.4 Strongly ionising radiation

Table 1 – High energy radiation energy range definition

Vacuum considerations

The magnitude of the absorption and scattering of useful x-ray / electrons signal by gas particles depends on the ‘quality’ of the vacuum of the experiment environment.

Vacuum regime Pressure range[2]
mbar bar Pascal (Pa) Torr
Atmospheric pressure 1.013x10+3 1.013 1.013x10+5 7.6x10+2
High vacuum 1x10-3 to 1x10-9 1x10-6 to 1x10-12 1×10−1 to 1×10−7 7.5x10-4 to 7.5x10-10
Ultra high vacuum 1x10-9 to 1x10-12 1x10-12 to 1x10-15 1×10−7 to 1×10−10 7.5x10-10 to 7.5x10-13
Extremely high vacuum <1x10-12 <1x10-15 <1×10−10 <1×10M−13

Table 2 – Vacuum regime definition

Vacuum chambers detector coupling considerations

Examples of vacuum environments onto which Andor ‘open front’ ‘SO’ detectors can be attached are shown on fig. 1.

test

Vacuum Chamber

Example of vacuum assemblies accepting Andor ‘SO’ detectors

Vacuum Ultraviolet (VUV) spectrograph

Example of vacuum assemblies accepting Andor ‘SO’ detectors

Extreme Ultraviolet (EUV/XUV) spectrograph

Figure 1 – Example of vacuum assemblies accepting Andor ‘SO’ detectors

Material considerations and bake-out

Materials for use in vacuum should:

  • Exhibit low rate of outgassing (release of adsorbed gas or water molecules trapped in the material)
  • Be carefully machined to minimize the amount of cracks and/or defects that could trap unwanted gas or moisture
  • Be tolerant to bake-out temperatures

A bake-out process is used to force impurities out of the surfaces exposed to the vacuum, as these could cause outgas and impact on the quality of the vacuum. This could in turn affect the detector minimum cooling temperatures.

Andor vacuum flanges are machined from stainless steel 304 grade.

Flange standards considerations

There are 3 main flange standards used to couple cameras to vacuum enclosures. These formats are described in the following sections:

  1. CF (Conflat™) standard
  2. ISO standard (including ISO-F and ISO-K)
  3. KF standard

Other flange standards, such as Japanese ‘JIS’[3] or Korean ‘KS’, can be considered through Andor's Customer Special Request (CSR) process. Please contact your local Andor representative for further information.

Connecting to VUV/EUV/XUV spectrographs

Andor detectors can be integrated on a number of 3rd party VUV, EUV or XUV spectrographs, including McPherson[4] or H+P Spectroscopy[5]. However these spectrographs may not be designed with the connection standards listed above.

Please refer to the McPherson spectrograph section of this document for further details on available options. For any other option, please contact your local Andor representative for further information.

1. CF (or ConFlatTM) / ICF format

This flange format is based on a metal-to-metal sealing by mean of a knife-edge interface. It is suitable for operation in environments ranging from atmospheric pressure to ultrahigh vacuum (< 10−13 Torr or < 1.3x10-13 mbar) - Researchgrade ‘direct detection’, open-front cameras are compatible with pressure environments down to 10-8 / 10-9 mbars.

Typical CF flange interface, with soft metal gasket

Figure 2 – Typical CF flange interface, with soft metal gasket shown on the drawing on the right

A knife-edge sealing involves a sharp-edged wedge of hard material on one side of the flange that cuts into a soft metal gasket on the mating part as the attachment bolts are tightened. The ductility of the soft metal (e.g. oxygen-free copper) gasket is used to fill small defects in the flange, providing an extremely leak-tight seal while also compensating for potential thermal expansion mismatch during bake-out up to ~ 450°C. Note that detectors cannot be exposed to such temperature – the maximum sustainable bake-out temperature for Andor cameras is +55°C.

Two different definition standards exist to specify this flange formats – equivalences are shown in table 3:

  • European/Asian standard uses the letters ‘DN’ followed by the inner diameter (ID) of the flange in mm, e.g. DN10 denotes a flange with 10 mm internal diameter.
  • American standard uses a number to define the outer diameter (OD) of the flange in inches.
Europe, Asia [ID, mm] America [OD, inches] Notes
DN10CF 1 (1)
DN16CF 1⅓ ("mini") (1)
DN25CF 2⅛ (1)
DN40CF 2¾ (1)
DN50CF 3⅜ (1)
DN50CF 4½ (2)
DN75CF 4⅝ (1)
DN100CF 6 (3)
DN125CF 6¾ (5)
DN160CF 8 (4) (5)
DN200CF 10 (5) (6)
DN250CF 12 (5) (6)

Table 3 – Equivalence of common CF flange formats

(1) Not suitable for Andor cameras | (2) Suitable for Newton and iKon-M | (3) Andor standard CF configuration for Newton, iKon- M and iKon-L platforms | (4) Andor standard CF configuration for iKon-XL | (5) These I.D. can also be accommodated for Newton, iKon-M and iKon-L. Please contact your local Andor representative for further information | (6) These I.D. can also be accommodated for iKon-XL. Please contact your local Andor representative for further information.

The following table summarises Andor’s standard flange interface technical characteristics. Please contact your local Andor representative for alternative formats:

Camera platform (a) Newton DO920P-xx iKon-M DO934P-xx iKon-L DO936N-*0W-xx (c) iKon-XL XLO-EAyy-C0
Vacuum chamber standards compatibility DN100CF / 6” CF / CF-152 DN100CF / 6” CF / CF-152 DN100CF / 6” CF / CF-152 DN160CF / 8” CF / CF-203
Outer diameter (OD) (b) 5.96” [151.6 mm] 5.96” [151.6 mm] 5.96” [151.6 mm] 7.97” [202.4 mm]
Inner diameter (ID) (b) 1.97” [50.0 mm] 1.97” [50.0 mm] 2.68“ [68.0 mm] 4.41” [112 mm]
Knife-edge diameter (KD) 4.52” [114.9 mm] 4.52” [114.9 mm] 4.54“ [115.3 mm] 6.525” [165.7 mm]
Mounting holes PCD 16 x ø8.5 mm thru on PCD ø130.3 mm 16 x ø8.5 mm thru on PCD ø130.3 mm 16 x threaded on PCD ø130.3 mm 20 x ø8.5 mm thru on PCD ø181.1 mm
Mounting holes format Clearance for M8 or 5/16 UNC bolts Clearance for M8 or 5/16 UNC bolts Threaded holes (c) Clearance for M8 or 5/16 UNC bolts
Bolt mounting direction From camera to vacuum chamber From camera to vacuum chamber From vacuum chamber to camera From camera to vacuum chamber
Flange orientation Rotatable (11.25º increments) Rotatable (11.25º increments) Fixed (mounting hole at 12 o’clock) Rotatable (11.25º increments)
Flange to sensor plane distance (nominal) 0.13“ [3.2mm] 0.13“ [3.2mm] 0.24” [6.1mm] 0.24” [6.1mm]
Filter holder type Bayonet (d) Bayonet (d) Screw-in (e) Screw-in (e)

Table 4 – Andor standard CF interfaces definition

(a) xx = sensor code, e.g. BN, BR-DD; yy = sensor code identifier (iKon-XL only) | (b) The O.D. is used to describe the camera flange, while the I.D. is specified according to the sensor size | (c) There are 3 specific threaded hole options for iKon-L: a) M8 thread built into the flange - camera code *= M0W, b) 5/16 UNC thread built in to flange - camera code *= I0W, c) Encapsulated nut (either M8 or 5/16UNC) – camera code *= 00W [nuts can be removed and replaced if thread is worn] | d) Included with camera | (e) Optional

Details of flange characteristic dimensions

Figure 3 – Details of flange characteristic dimensions

2. ISO format (ISO-K and ISO-F)

This flange format is based on an O-ring seal. It is suitable for operation in environments ranging from atmospheric pressure to high vacuum (10−8 Torr or 1.3x10-8 mbar). - Research-grade ‘direct detection’, open-front cameras can work with pressure environments down to 10-6 / 10-7 mbars. The ISO format includes two main variations:

  • ISO-F, using conventional bolt/nuts to secure the flange mating parts
  • ISO-K, using claw clamps to secure the flange mating parts
ISO-F and ISO-K flange interface with assembled and exploded view

Figure 4 – ISO-F (top) and ISO-K (bottom) flange interface with assembled (left) and exploded view (right)

The O-ring assembly consists of a centring metal ring and an O-ring wrapped around it. The two flange mating parts are designed with a groove that is used to centre the O-ring assembly while tightening the attachment bolts or clamps.

This interface can sustain bake-out temperatures up to ~ 150oC. Note that detectors cannot be exposed to such temperature – the maximum sustainable bake-out temperature for Andor cameras is +55oC.

ISO flange formats are identified by the nominal internal diameter (I.D.) in millimetres, e.g. an ISO-K 100 or ISO-F 100 have an internal diameter of 100 mm. The standard sizes range from ISO-63 to ISO-630[6].

ISO-K 100 or ISO-F 100 are recommended for Newton and iKon-M series, but can also be provided for iKon-L. Other larger O.D. can however be accommodated. Please contact your local Andor representative for further information.

3. KF format

This flange format – also known as Klein Flange (or KF) - is also based on an O-ring seal. It is suitable for operation in environments ranging from atmospheric pressure to high vacuum (10−8 Torr or 1.3x10-8 mbar).

KF flange interface with assembled and exploded view

Figure 5 – KF flange interface with assembled (left) and exploded view (right)

The O-ring assembly consists of a centring metal ring and an O-ring wrapped around it. The two mating parts of the flange are designed with a chamfered lip and an interface to locate the O-ring assembly. A circular claw with a wingnut (or thumbscrew) is used to compress the O-ring assembly and the two flange parts to create the seal.

This interface can sustain bake-out temperatures up to ~ 150°C. Note that detectors cannot be exposed to such temperature – the maximum sustainable bake-out temperature for Andor cameras is +55°C.

KF flange format is identified by the nominal internal diameter (I.D.) in millimetres, e.g. an KF100 has an internal diameter of 100 mm. The standard sizes range from KF10 to KF50[6].

Please contact your local Andor representative for further information on options available.

4. McPherson spectrographs

Andor Newton CCD detectors options -995 can connect readily to the spectrographs listed in table 5.

The Newton -995 cameras (e.g. DO920P-BN-995 or DO940P-BN-995) coupling interface is an O-ring sealed flange, designed with the sensor image plane sitting 5.4 mm behind the faceplate front and with a 38 x 25mm aperture. Please refer to[7] for software control of these spectrographs, and see appendix A for further details on the cameras.

Model number Wavelength Range (nm) Focal length (mm) Spectral resolution (nm) Multi-track High vacuum Optical design
234/302 VUV 30 to 550 200 0.1 - Yes Concave Holographic
225 VUV 30 to 1,200 1,000 0.015 Yes Yes Normal Incidence
248/310G Soft X-ray EUV, XUV < 1 to 310 1,000 0.018 - Yes Grazing Incidence

Table 5 – Main VUV/EUV/XUV spectrographs compatible with Andor -995 flange variation

Filters

Andor open-front ‘SO’ cameras can be coupled to filters to remove the unwanted contribution of low energy radiations on the useful signal.

For x-ray experiments with useful information in the energy range 3-30 keV (‘direct’ x-ray detection with a siliconbased detector), thin foils of Beryllium (Be) or Aluminium (Al) are typically used. Typical transmission characteristics of Beryllium filters as a function of thickness are shown on fig. 6.

Typical Beryllium filters transmission curves versus material thickness

Figure 6 – Typical Beryllium filters transmission curves versus material thickness

Camera platform Standard Be filter thickness Standard Be filter diameter [clear aperture] Min. recommended thickness *
Newton SO 250 μm Ø45.5 ±0.2 mm [Ø35.0mm] 200 μm
iKon-M SO 250 μm Ø45.5 ±0.2 mm [Ø35.0mm] 200 μm
iKon-L SO 250 μm Ø60.0 ±0.2 mm [Ø45.0mm] 250 μm
iKon-XL SO 250 μm Ø105.0 ±0.2 mm [Ø95.0mm] 250 μm

Table 6 – Standard filter thickness for Andor open-front ‘SO’ platforms
* available through CSR – thinner windows carry increased risks of breakage, long-term stress

Other filter options[8] may be considered through Andor’s CSR process. Please contact your local Andor representative for further information. See Appendix B for further information on filters and filter holders.

There are two options for coupling filters to Andor open-front ‘SO’ cameras:

  1. A vacuum chamber-compatible filter holder than can be attached to the camera through a bayonet or a screw-in interface (depending on the camera interface, see table 4)
  2. A filter mount that allows the camera to be decoupled from a vacuum chamber and be used as a standalone detector (or ‘SY’). Note that this configuration cannot be attached to another vacuum chamber

Vacuum chamber-compatible filter holder

Newton and iKon-M (top) and iKon-L (bottom) vacuum chamber-compatible filter holders

Figure 7 – Newton and iKon-M (top) and iKon-L (bottom) vacuum chamber-compatible filter holders

The filter holders SO-FILTER-HOLDER-xx have a built-in pumping channel to ensure pressure equilibrium on either side of the filter, hence minimizing the risk of mechanical stress and damage to the filter.

Standalone camera filter holder (SO-to-SY camera type converter)

SO-to-SY filter holder assembly for Newton and iKon-M and iKon-L open-front ‘SO’ platforms

Figure 8 – SO-to-SY filter holder assembly for Newton & iKon-M (top) and iKon-L (bottom) open-front ‘SO’ platforms

FLG-SO-SY-CONVERT-45 mounts onto the Newton and iKon-M ‘SO’ camera series. FLG-SO-SY-CONVERT-60 mounts onto the iKon-L ‘SO’ camera series. The vacuum sealing interface is a knife-edge.

The vacuum between the filter and the sensor is achieved by connecting a pump to the interface highlighted on fig. 7, which is fitted with a KF16 interface. Note that since this interface relies on an O-ring vacuum seal, pressure down to 10-6 mbar can be achieved. Continuous pumping is recommended to maintain the highest degree of vacuum, and also ensure access to the lowest camera cooling temperature.

Appendix A – Andor ‘open-front’ ‘SO’ detectors for VUV / EUV / XUV imaging and spectroscopy

Newton CCD platform and iKon-M CCD platform

Newton CCD platform and iKon-M CCD platform

Sensor quantum efficiency (QE) considerations

Test

Newton CCD platform and iKon-M CCD platform

Sensor type key:

BN: Back-illuminated un-coated sensor
BR-DD: Back-illuminated AR coated deep-depletion sensor
FI: front-illuminated

Note that other sensors could be considered for integration into Andor ‘open-front’ ‘SO’ platforms – please contact your local Andor representative for further details

BN-DD: Back-illuminated deep-depletion un-coated sensor – increased higher energy QE versus -BN type
FI-DD: Front-illuminated deep-depletion sensor – increased higher energy QE versus -FI type

Detectors for Spectroscopy

  Newton DO920P-XX Newton DO920P-BN-995 Newton DO940P-XX Newton DO940P-BN-995
Sensor format 1024 x 255 matrix
26 μm pixels
e2v CCD 30-11
2048 x 512 matrix
13.5 μm pixels
e2v CCD 42-10
Sensor Size 26.7 x 6.7 mm 27.6 x 6.9 mm
Sensor options (-XX) BEN, BR-DD, FI BEN BEN, BN, FI BEN, BN
Minimum TE-cooling temperature -100 °C -100 °C -100 °C -100 °C
Maximum spectral rate 273 sps (>1,580 with crop mode) 122 sps (>940 with crop mode)
Pixel well depth 500,000 e- 100,000 e-
Register well depth 1,000,000 e- 150,000 e- (HS mode)
600,000 e- (HC mode)
Readout noise 4 e- @ 50 kHz readout 3.5 e- @ 50 kHz readout
PC interface USB 2.0
Flange interface DN100CF / 6” CF / CF-152 O-ring seal, sensor 5.4 mm behind faceplate, 38 x 25 mm aperture DN100CF / 6” CF / CF-152 O-ring seal, sensor 5.4 mm behind faceplate, 38 x 25 mm aperture

Detectors for Imaging

Detectors for imaging iKon-M DO934P-XX iKon-L DO936N-*0W-XX iKon-XL 230 XLO-EAxx-C0 iKon-XL 231 XLO-EAxx-C0
Sensor format 1024 x 1024 matrix
13 μm pixels
e2v CCD 47-10
2048 x 2048 matrix
13.5 μm pixels
e2v CCD 42-40
e2v CCD 230-84
4096 (H) x 4112 (V) matrix 15 μm pixels
e2v CCD 231-84
4096 (H) x 4112 (V) matrix 15 μm pixels
Sensor Size 13.3 x 13.3 mm 27.6 x 27.6 mm 61.4 x 61.7 mm
Sensor options (-XX) BEN, BN, BR-DD, FI BEN, BN, BR-DD, FI BN BEN, BN, BR-DD
Minimum TE-cooling temperature -100 °C -100 °C -75 °C
Maximum image rate 4.4 fps 0.95 fps >0.5 fps >0.35 fps
Pixel well depth 100,000 e- 100,000 e-
150,000 e- [BR-DD]
150,000 e- 350,000 e-
Output node capacity well depth 250,000 e- 1,000,000 e- 450,000 e- [HS]
900,000 e- [HC]
600,000 e-
Readout noise 2.9 e- @ 50 kHz
3.3 e- @ 50 kHz [BR-DD]
2.9 e- @ 50 kHz
4.3 e- @ 50 kHz
4.5 e- @ 100 kHz 2.1 e- @ 100 kHz
PC interface USB 2.0 USB3.0, fibre-optic
Flange interface DN100CF / 6” CF / CF-152 DN100CF / 6” CF / CF-152 DN160CF / 8” CF / CF-203

Table 7 – Andor ‘Open-front’ cameras key specifications

Appendix B – ‘Open-front’ ‘SO’ cameras accessories

Camera platform Filter holder part number Standard Be filter dimensions (Ø x thickness) Standard filter part number (order seperatly) Min. recommended thickness * Max. recommended thickness
Newton Included with camera Ø45.5 mm x 250 μm ACC-OPT-02839 200 μm 500 μm
iKon-M Included with camera Ø45.5 mm x 250 μm ACC-OPT-02839 200 μm 500 μm
iKon-L SO-FILTER-MNT-IKONL Ø60.0 mm x 250 μm ACC-OPT-03838 250 μm 500 μm
iKon-XL SO-FILTER-MNT-IKONXL Ø105.0 mm x 250 μm ACC-OPT-10395 250 μm 500 μm

Table 8 – Standard filter and holder references for Andor open-front ‘SO’ platforms
* available through CSR – thinner windows carry increased risks of breakage, long-term stress

Note: Copper gasket for CF ConflatTM flanges can be sourced from a number of vendors. This part can also be ordered from Andor under the reference ACC-FLG-SO-GSKT-CU.

Appendix C – Q&A

Q1: Does the vacuum chamber pressure impact the minimum cooling temperature achievable by an ‘openfront’ ‘SO’ camera?

A1: No impact on cooling temperatures for pressure below ~10-5 mbar

Q2: What information should be provided for custom ConflatTM flange design?[6]

Parameter Value Notes
Vacuum chamber standards compatibility?   e.g. DN200CF or 12” (if applicable)
Outer diameter? mm or inches Ø45.5 mm x 250 μm
Inner diameter? mm or inches this can refer to the clear aperture around the sensor
Knife-edge diameter? mm or inches  
Mounting holes pitch circle diameter (PCD)? mm or inches  
Mounting hole format? mm or inches e.g. M8 or 3⁄8
Bolts mounting direction?   this is the direction from which the screws should inserted from, e.g. camera flange-to-vacuum chamber, or vacuum chamber-to-camera flange
Flange rotation?   e.g. Yes/No, step granularity (in degrees)
Flange to sensor plane distance (nominal) mm or inches from the front face of the camera to the sensor plane
Filter holder requirement   e.g. Yes / No, filter type / material / thickness

Q3: Can Andor ‘SO’ cameras accommodate existing vacuum chamber setups designed with ‘straddled’ flange configuration?

Platform 6” flange ‘straddled’ configuration 4.5” flange ‘straddled’ configuration
Newton ‘SO’ ? YES
Newton ‘SO’ models can accommodate 11.25º step rotation as standard
Flange to sensor plane distance = 0.13“ [3.2mm]
Compatible with standard Newton ‘SO’ bayonet filter holder configuration [included]
! More information required
Please contact your local Andor representative do discuss your specific VUV/EUV/XUV spectrograph focal plane configuration
iKon-M ‘SO’ ? YES
iKon-M ‘SO’ models can accommodate 11.25º step rotation as standard
Flange to sensor plane distance = 0.13“ [3.2mm]
Include standard iKon-M ‘SO’ bayonet filter holder
? YES
Variation ‘-9NP’ (e.g. DO934P-BN-9NP) can accommodate straddled configuration
Flange to sensor plane distance = 0.23“ [5.9 mm]
Include bayonet filter holder
iKon-L ‘SO’ ? YES
Variation ‘-9OW’ (e.g. DO936N-*0W-BN-9OW) can accommodate straddled configuration
Flange to sensor plane distance = 0.24“ [6.1 mm]
Include standard iKon-L ‘SO’ screw-in filter holder
Not available
Illustration of vacuum flanges with ‘straddled’ configuration (left) and Andor standard configuration (right)

Figure 10 – Illustration of vacuum flanges with ‘straddled’ configuration (left) and Andor standard configuration (right)

Q4: What information should be provided for ISO flange design?[6]

Parameter Value Notes
Vacuum chamber standard compatibility (if any)?   e.g. ISO-F 100 or ISO-K 100
Tube outer diameter? mm or inches  
Flange outer diameter? mm or inches  
Flange thickness? mm or inches  
Number of bolts (ISO-F)?   e.g. 8x Ø9 mm on Ø145 mm
Centring O-ring assembly characteristics?   e.g. ISO 100 format, or alternatively define the lip internal and external diameter (mm or inches) and the thickness (mm or inches)
Bolts mounting direction?   this is the direction from which the screws should inserted from, e.g. camera flange-to-vacuum chamber, or vacuum chamber-to-camera flange
Flange rotation?   e.g. Yes/No, step granularity (in degrees)
Flange to sensor plane distance (nominal) mm or inches from the front face of the camera to the sensor plane

Appendix D - Useful documents and links

  1. ISO 21348 Definitions of Solar Irradiance Spectral Categories, http://www.spacewx.com/pdf/SET_21348_2004.pdf
  2. Pressure conversion tool http://www.unit-conversion.info/pressure.html
  3. Japanese flange standard JIS: http://www.baleng.com.au/flanges-japanese-standard.php
  4. McPherson Vacuum Ultraviolet Soft X-ray & Extreme Ultraviolet spectrographs http://www.mcphersoninc.com/spectrometers/xuhvvuvuv/xuhvvuvuv.html andhttp://www.mcphersoninc.com/spectrometers/vuvuvvis/vuvuvvis.html
  5. H+P flat-field XUV / EUV / VUV spectrographss http://www.hoerlein-partner.com
  6. Lesker website http://www.lesker.com/newweb/menu_flanges.cfm
  7. Andor software control for 3rd party spectrograph technical note available at MyAndor
  8. Beryllium and Aluminium filters transmission http://henke.lbl.gov/optical_constants/filter2.html

Other useful references

  1. Andor High Energy Detection product brochure https://andor.oxinst.com/assets/uploads/Andor_High_Energy_Detection_Brochure.pdf
  2. Tips for CF vacuum flange coupling / installation http://www.lesker.com/newweb/flanges/flanges_technicalnotes_conflat_1.cfm#CF_Installation

← Back to Learning