Laser Specifications

Output Specifications
Single VLE maximum output power ≤ 500 mW Maximum of 2 lasers can be turned on at once under Fusion control. This is an IEC 60825-1 classification value and is far greater than in normal use. See the ‘specifications for the individual lasers on the following pages and more information in
Dual-VLE BCU maximum output power ≤ 500 mW Maximum of 4 lasers can be turned on at once under Fusion control. See previous note.
Operating wavelength range 400 to 800 nm  
Output Fibre
Fibre type Multimode (MM)  
Connector type FC/APC  
Beam Divergence
Beam divergence at internal module output 0.5-3 mrad (full angle) The divergence depends on the properties of the laser inside the VLE.
Beam divergence at VLE coupler output 0.025-0.13 NA (3-15 degrees full angle)
Beam divergence at VLE 50 μm fibre output ≤0.12 NA (14 degrees full angle)  
Beam divergence at VLE 25 μm fibre output ≤0.10 NA (12 degrees full angle)  
Beam divergence at BCU fibre output ≤0.14 NA (16 degrees full angle)  
Beam Diameter
Beam diameter at internal module output 0.3 - 3 mm The diameter depends on the properties of the laser inside the VLE
Beam diameter at VLE coupler output 20 - 100 μm
Beam diameter at VLE 50 μm fibre output 50 μm  
Beam diameter at VLE25 μm fibre output 25 μm  
Beam size at BCU fibre output 175 x 175 μm Fibre has square profile, so beam is square at fibre output

VLE Laser Output Power Specifications

    Power at internal module output, mW Power at coupler output, mW Power at 50 μm, 0.12 NA, fibre output, mW
Order code Laser wavelength, nm Typical Maximum Typical Maximum Typical Maximum
LM-405-488-DUAL-VLE 405 200 215 170 190 150 165
LM-445-515-DUAL-VLE 448 215 240 190 210 160 175
LM-405-488-DUAL-VLE 488 160 180 140 155 120 130
LM-445-515-DUAL-VLE 515 155 170 140 155 120 130

LM-561-50-VLE

or

LM-561-50-VLESLT3

561 45 55 40 45 35 45

LM-561-100-VLE

or

LM-561-100-VLESLT3

561 95 105 85 95 75 85

LM-561-150SL-VLE

or

LM-561-150SL-VLESLT3

561 140 160 130 140 110 120

LM-594-100SL-VLE

594 100 110 90 100 80 90

LM-640-730-DUAL-VLE

or

LM-640-VLE

638 195 215 165 185 140 155

LM-685-780-DUAL-VLE

or

LM-685-780-DUAL-VLE-SLT2

685 125 140 105 120 75 85
LM-640-730-DUAL-VLE 730 185 200 135 150 95 105

LM-685-780-DUAL-VLE

or

LM-685-780-DUAL-VLE-SLT2

780 210 230 180 195 125 140
Wavelength tolerances: 405 +5/-7 nm; 445 +7/-3 nm; 561 & 594 ±2 nm; 685 & 780 ±3 nm; 488, 515, 638, 730 ±5 nm;

BCU Laser Output Power Specifications

    Power at fibre output for BCU-120 and BCU-220, mW
Order code Laser wavelength, nm Typical Maximum
LM-405-488-DUAL-VLE 405 90 100
LM-445-515-DUAL-VLE 448 100 110
LM-405-488-DUAL-VLE 488 80 90
LM-445-515-DUAL-VLE 515 75 85

LM-561-50-VLE

or

LM-561-50-VLESLT3

  561 25 30

LM-561-100SL-VLE

or

LM-561-100SL-VLESLT3

55 60

LM-561-150SL-VLE

or

LM-561-150SL-VLESLT3

80 85

LM-594-100SL-VLE

594 60 65

LM-640-730-DUAL-VLE

or

LM-640-VLE

638 100 110

LM-685-780-DUAL-VLE

or

LM-685-780-DUAL-VLE-SLT2

685 45 55
LM-640-730-DUAL-VLE 730 65 75

LM-685-780-DUAL-VLE

or

LM-685-780-DUAL-VLE-SLT2

780 80 90
Wavelength tolerances: 405 +5/-7 nm; 445 +7/-3 nm; 561 & 594 ±2 nm; 685 & 780 ±3 nm; 488, 515, 638, 730 ±5 nm;