Electrical Power Specifications

  High Power Laser Engine
Mains input for main HLE unit 100 – 240 VAC, 50 – 60 Hz
Current rating for main HLE unit* 3 A
Mains overvoltage category CAT II An overvoltage category of CAT II means that the equipment is designed to cope with transient voltages above the rated supply that would be experienced by any product connected to a standard single-phase mains socket in a building.

*Current rating includes what HLE main unit is capable of for future purposes. This will therefore give a power consumption figure which is greater than what the current laser configurations will consume.

  Power Consumption, W
HLE with 2 diode lasers on and 560 laser on 70 Typical/140 Max.

HLE chassis

HLE key switch off (heaters stay on)

HLE on (fans off), lasers off

HLE on (fans on), lasers of

 

23 Typ.

25 Typ.

44 Typ.

HLE lasers/modules

LM-405-488-DUAL-MM

LM-445-515-DUAL-MM

LM-640-730-DUAL-MM

LM-HLE-560-1000 (HLE-FL1)

HLE-FL1 External Power Supply (PS-13)

LM-HLE-561-150/ LM-HLE-594-100 (OBIS LS)

 

21 Typ.

25 Typ.

11 Typ.

25 Typ. (excluding PS-13)

10 Typ.

8 Typ./13 Max.

BCU Power Consumption

BCU + External Power Supply: 1.5 W typical

BCU Only: 1 W typical