Polybutadiene (or rubber) is added during polymerisation of styrene to increase the flexibility of polystyrene polymers. It is important to closely monitor the amount of polybutadiene added, so that the flexibility of the polystyrene polymer may be regulated. NMR can provide an effective and rapid method for the regulation of polybutadiene content in Impact Polystyrene.
NMR has a number of advantages over other techniques:
12 samples of impact polystyrene were used for the calibration of polybutadiene content. Each sample was weighed into a tared 26mm glass tube before measurement at room temperature in the MQC+. The resulting calibration graph is shown in Figure 1.

Fig. 1 — Calibration graph of NMR signal per unit mass against polybutadiene content obtained by a reference method. Correlation Coefficient r = 1, Standard Deviation = 0.15.
Table 1 shows that the repeatability (or precision) of the NMR data is excellent.
| Repeat number | Polybutadiene Content (%) |
| 1 | 6.79 |
| 2 | 6.81 |
| 3 | 6.80 |
| 4 | 6.81 |
| 5 | 6.79 |
| 6 | 6.78 |
| 7 | 6.81 |
| 8 | 6.79 |
| 9 | 6.79 |
| 10 | 6.80 |
| Average | 6.80 |
| Standard Deviation | 0.01 |
Table 1: NMR measurement repeatability.
The MQC+ with a 0.55 Tesla (23 MHz hydrogen resonance) magnet, fitted with a 26mm diameter (14ml sample) probe is a suitable instrument for this application. The Rubber in Impact Polystyrene package consists of:
In addition to this package you may also require:
The instrument offers advantages over others on the market:
Note: Other instruments/packages are available for the analysis of larger or smaller quantities of sample. Please contact Oxford Instruments for details.