Cocoa beans are processed to extract cocoa liquor and cocoa butter, both major ingredients of dark and milk chocolate. Cocoa beans are roasted, separated from their shell and cracked into nib. Nib is then ground to produce cocoa liquor or cocoa butter plus cocoa powder. As the quality of the beans can vary depending on the environmental conditions in the region where they were grown, it is important to quantify the fat content of the raw and intermediate materials to ensure consistency of the final product.
The International Office of Cocoa, Chocolate and Sugar Confectionery Industries (IOCCC) 14-1972 and AOAC 963.15 methods involve acid digestion to release the bound lipids followed by extraction using petroleum ether. However, these methods are time consuming, require skilled operators and the use of hazardous solvents.
The MQC+ benchtop Nuclear Magnetic Resonance (NMR) analyser provides an alternative method to wet chemistry; it is quick and easy to perform, simple to calibrate, and requires minimal sample preparation. As such it is ideal for routine operation without any requirement for hazardous chemicals or specialist operator training.
The analytical technique is based on direct measurements of the NMR response obtained from the fat in chocolate products normalised by the sample mass. The fat content is quantified by calibrating the NMR signal per gram of sample against concentration determined by a suitable reference technique.
Ultimately, only two well known standards are required to calibrate the instrument as the calibration is linear. However initially it is recommended that the instrument is calibrated by 3–6, preferably more, standards with known fat contents evenly spread over the range of interest.
NMR is a comparative technique therefore cannot be more accurate than the reference technique against which it is being compared. However, it is more robust and reproducible than extraction methods and therefore calibration error is reduced by analysing more reference samples.
The melted chocolate is poured into a pre-tared vial and weighed. The vial is then placed in a temperature controlled conditioning block at 50°C for 20 minutes prior to analysis. Inserting the sample into the instrument automatically starts the NMR analysis after which it displays the fat content after 8 seconds.
Calibrations were obtained by measuring various samples of dark and milk chocolate (Figure 1) and chocolate crumb (Figure 2), a mixture of milk, sugar and cocoa liquor, and plotting the NMR data against the fat contents obtained by the International Office of Cocoa Chocolate and Sugar Confectionery (IOCCC) reference method.
The results for both chocolate and crumb show excellent correlations with R values of 0.999 and 0.997 respectively. The reference values are compared against those calculated from the NMR calibration in Tables 1 and 2.

Fig. 1 — NMR calibration for Total Fat Content of dark and milk chocolate referenced against the IOCCC method. The correlation coefficient (R) and standard deviation (σ) are 0.999 and 0.07 respectively.

Fig. 2 — NMR calibration for Total Fat Content of chocolate crumb referenced against the IOCCC method. The correlation coefficient (R) and standard deviation (σ) are 0.997 and 0.12 respectively.
| Sample | Ref. Value (%) | NMR Value (%)* | Diff. (%) |
| 1 | 29.7 | 29.7 | 0.0 |
| 2 | 28.6 | 28.7 | 0.1 |
| 3 | 33.5 | 33.6 | -0.1 |
| 4 | 27.5 | 27.5 | 0.0 |
| 5 | 26.1 | 26.1 | 0.0 |
| 6 | 31.6 | 31.5 | -0.1 |
| 7 | 28.8 | 28.6 | -0.2 |
Table 1. A comparison of the fat contents of various samples of dark and milk chocolate obtained from the reference technique against those calculated from the NMR calibration in Figure 1. *Average of two subsamples.
| Sample | Ref. Value (%) | NMR Value (%)* | Diff. (%) |
| 1 | 14.5 | 14.5 | 0.0 |
| 2 | 15.3 | 15.3 | 0.0 |
| 3 | 15.9 | 15.8 | -0.1 |
| 4 | 16.9 | 16.9 | 0.0 |
| 5 | 16.5 | 16.6 | 0.1 |
| 6 | 17.2 | 17.4 | 0.2 |
| 7 | 19.7 | 19.7 | 0.1 |
Table 2. A comparison of the fat contents of various samples of chocolate crumb obtained from the reference technique against those calculated from the NMR calibration in Figure 2. *Average of two subsamples.
Oxford Instruments offers a package especially tailored to the measurement of Total Fat Content in Chocolate and other Cocoa Derivatives:
In addition to the above, the following are also required: