Fish feed manufacturers produce pellets that vary in size, oil content (from 5 to 40%) and moisture (usually less than 10%). Measuring the oil and moisture content is essential for quality control to ensure that the products meet their nutritional values. The amount of oil used is particularly important because of the significant cost of the raw material and the affect it may have on the texture and perceived quality of the product. Low resolution Nuclear Magnetic Resonance (NMR) provides a method which is quick and easy to perform, simple to calibrate, and not dependent on the sample matrix.

Solvent extraction techniques are commonly used for determination of fat content. However, they tend to be slow, inaccurate and require skilled personnel. In addition, many of the often hazardous chemicals used are becoming increasingly unacceptable according to international environmental standards. Despite these issues solvent extraction continues to be used as a reference measurement for quality control.
Instrumental methods are often referred to as secondary techniques since they are usually set up to match the results produced by solvent extraction. To provide a result equivalent to the traditional extraction techniques, secondary techniques require a correlation against the reference technique used. Although they are fast and easy to maintain, many secondary techniques need to be calibrated and maintained regularly. Also, maintenance and consumables add significantly to the cost of ownership.
For example, although Supercritical Fluid Extraction (SFE) is reasonably fast, it requires high maintenance and the cost of compressed CO₂ used to extract oil is also significant. Near Infra-Red (NIR) is usually used for trend analysis and is often applied to on-line monitoring. However it is difficult to apply on opaque samples as it can only scan the surface. It is also complex to calibrate and difficult to maintain accurate results on a large variety of product types. For that primary reason, NIR has limited applicability for the quality control of oil content in fish feed of various types.
NMR has been used for decades to measure oil and moisture contents (for instance, the ISO10565 standard method for seeds) which has also been found to be applicable to fish feed. NMR has great advantages over other secondary techniques.
Samples are simply loaded into pre-tared glass vials, weighed, conditioned, then inserted into the instrument which detects the sample, automatically starting the NMR analysis. The instrument returns the oil and moisture content values in less than one minute.
*It is recommended to pre-condition the sample for 15 minutes at 40°C for better precision.
The NMR instrument is initially calibrated using a 100% oil sample or a set of reference standards with known oil and moisture contents. Several samples of fish feed were analysed by NMR and their oil and moisture values predicted by the calibration were compared against the contents determined by solvent extraction and oven drying respectively.
Figures 1 and 2 show those results align on a straight line demonstrating the accuracy of the NMR technique. The measurement precision is typically less than 0.2%.

Figure 1 — Comparative results between predicted NMR results and reference oil contents (%)

Figure 2 — Comparative results between predicted NMR results and reference moisture contents (%)
The MQC+ with a 0.55 Tesla magnet fitted with a 26 mm diameter (10ml) probe is ideal for this application. The "Oil and Moisture in Fish Feed" package comprises: