When the first commercial NMR instruments were introduced in the early 1950s, they could analyse a single nucleus: hydrogen-1 (1H, proton). Only many years later would commercial multinuclear spectrometers be introduced. They soon became a mainstay of analysis in both academia and industry, fuelling discoveries in fields ranging from chemistry to biotechnology, pharmaceuticals, polymers, and energy, among many others.
Benchtop NMR instruments, which offer NMR capabilities in a more compact, low-maintenance form factor, have followed a similar, if faster, development path. Since the introduction of the first, single-nucleus spectrometers in 2009, two- and three-nucleus versions quickly became available. Just a decade later, 2019 saw the debut of the X-Pulse 60, the first broadband multinuclear benchtop NMR spectrometer, expanding the reach of benchtop instruments into areas previously reserved for traditional high-field spectrometers. The 2025 launch of the X-Pulse 90 further extended broadband benchtop NMR spectroscopy to a higher frequency, with increased sensitivity and improved signal dispersion.
As broadband capabilities became available in compact benchtop systems, X-nuclei NMR spectroscopy is accessible to a wider range of users across new academic and industrial fields. However, before starting, it is important to understand which nuclei can be observed, which factors influence their detectability, and how instrument configuration and setup affect measurement feasibility and data quality.
The following sections introduce the fundamentals to X-nuclei spectrometry and explain how to tune the NMR spectrometer to the nuclei of interest to achieve best possible results.
Many elements are observable by NMR but not every atom of each element is. Chemical elements differ by the number of protons in their nucleus, indicated by their atomic number. Each atom of an element has the same number of protons. However, atoms of the same element can have different numbers of neutrons in the nucleus, producing multiple isotopes each with a unique atomic mass. This varying composition of atomic nuclei alters their physical properties, affecting their suitability for NMR spectroscopy.

Figure 1 Periodic Table of the Elements
Some of the properties relevant to NMR spectroscopy of selected isotopes are summarised in Table 1.
Table 1 NMR properties of selected isotopes
| Isotope | Nuclear Spin, I | Natural Abuncance (%) | Larmor Frequency @1.4 T (MHz) | Larmor Frequency @2.1 T (MHz) | Relative Receptivity versus 13C |
| Hydrogen-1 (proton) | 1/2 | 99.99 | 59.70 | 90.45 | 5870 |
| Hydrogen-2 (deuterium) | 1 | 0.0115 | 9.16 | 13.88 | < 0.01 |
| Lithium-7 | 3/2 | 92.41 | 23.20 | 35.15 | 1590 |
| Boron-10 | 3 | 19.90 | 6.41 | 9.72 | 23.2 |
| Boron-11 | 3/2 | 80.10 | 19.15 | 29.02 | 777 |
| Carbon-13 | 1/2 | 1.07 | 15.01 | 22.74 | 1 |
| Nitrogen-14 | 1 | 99.63 | 4.31 | 6.54 | 5.9 |
| Nitrogen-15 | 1/2 | 0.37 | 6.05 | 9.17 | 0.02 |
| Fluorine-19 | 1/2 | 100 | 56.17 | 85.11 | 4900 |
| Silicon-29 | 1/2 | 4.68 | 11.86 | 17.97 | 2.16 |
| Phosphorus-31 | 1/2 | 100 | 24.17 | 36.61 | 391 |
| Cobalt-59 | 7/2 | 100 | 14.17 | 21.46 | 1640 |
| Platinum-195 | 1/2 | 33.83 | 12.83 | 19.44 | 20.7 |
Nuclear spin is one physical property relevant for NMR. In principle, all nuclei with a non-zero nuclear spin may be observed by NMR spectroscopy. In practice, other considerations, such as the abundance of the isotope, its sensitivity, and its typical linewidth, make some nuclei easier to analyse than others.
In nature, most chemical elements are found as a mixture of isotopes. The NMR signal is directly proportional to the number of nuclei in the sample producing that signal, therefore the abundance of an NMR active isotope greatly influences the ease of analysis.
For isotopes with low natural abundance, isotopic enrichment can increase sensitivity; doubling the percentage of a low abundance isotope in the sample would double the signal. One everyday example is the use of deuterated solvents, in which the percentage of deuterium (2H) is routinely increased from its natural abundance of 0.0115% to 99% or higher to provide a strong deuterium lock signal.
Another consideration is that each NMR active isotope has a different level of sensitivity, which is related to the frequency of the signal it produces. In a fixed magnetic field, each isotope resonates at a characteristic frequency (called the Larmor frequency).
Although the term "Larmor frequency" may be unfamiliar to casual NMR users, they probably use it every day, as magnet strengths are commonly expressed as the 1H Larmor frequency (for example, 90 MHz for the X-Pulse 90's 2.1 tesla permanent magnet). All other considerations being equal, in a fixed magnetic field, an isotope that produces a signal at a higher frequency will be more sensitive. Other than the extremely low-abundance, radioactive, tritium (3H) isotope; the high-abundance proton (1H) has the highest Larmor frequency, which is a major reason for its popularity in NMR.
The receptivity of an isotope quantifies the overall ease of observation by NMR spectroscopy, accounting for both its natural abundance and physical properties. Receptivity is usually reported either relative to proton or to carbon-13 (as done in Table 1). Comparing values of the different isotopes shows that observation of 1H is almost six thousand times "easier" than that of 13C.
Low receptivity in NMR spectroscopy can be overcome by increasing sample concentration, running longer experiments (increasing the number of scans), and choosing specific experiments designed to enhance sensitivity.
In addition to the properties of a nucleus, the capabilities of the NMR spectrometer are a critical consideration for X-nuclei NMR spectroscopy. To analyse a nucleus, the NMR spectrometer hardware must be designed to operate at its Larmor frequency.
Different nuclei have different Larmor frequencies. A proton, for example, will resonate at 90 MHz on a 2.1 tesla magnet, but 13C will resonate at just 22.7 MHz. As an example of the scope of an NMR spectrometer, the broadband X-channel of an X-Pulse 60 or 90 fitted with a standard broadband probe and preamplifier can analyse nuclei with Larmor frequencies between 41% and 19% that of the proton. A few of the nuclei in this range include 13C, 31P, 7Li, 11B, 23Na, 27Al, and 29Si.
Observing nuclei outside that range would require the probe and/or preamplifier to be exchanged for ones optimised for the frequence range of interest. For example, the low-frequence probe/preamplifier available for the X-Pulse 60 which will allow for the observation of nuclei with Larmor frequencies between 14% and 7% that of the proton (including: 2H, 15N & 14N).
To get the best performance out of an NMR spectrometer, it is important for the electrical circuit used to transmit and receive the radio-frequency (RF) signals to be tuned to the Larmor frequency of the nuclei, through a process known as tuning & matching (commonly referred to just as tuning).
Tuning & matching is performed by adjusting two capacitors within the probe. The first is used to tune the RF to the relevant nuclei and is analogous to tuning to a radio station; the second is used to match the impedance of the sample to the circuit. This ensures that the maximum possible RF energy is absorbed by the sample, and hence the strongest possible NMR signal is obtained.
Due to the wide frequency range of the broadband X-channel on the X-Pulse, it is also necessary to swap between various fixed capacitors to tune between the complete range of nuclei. While on high-field NMR spectrometers, these adjustments would usually be made on the bottom of the probe, so underneath the magnet; on an X-Pulse the various capacitors are easily accessible on the top of the magnet, next to the sample load position.

Figure 2 Oxford Instruments X-Pulse Broadband Benchtop NMR Spectrometer showing the variable and fixed capacitors for Tuning & Matching the X-channel.
An NMR spectrometer needs to be tuned correctly for a number of reasons. Not only will it maximise the sensitivity of the measurement; it ensures that the RF pulses are as short as possible and minimises the power required for decoupling. Most importantly it ensures that previously calibrated pulse widths (and powers) are reproducible, which is essential when performing advanced multi-pulse NMR experiments.
To tune & match an NMR spectrometer the frequency is swept over a narrow region centred on the target frequency, with the response compared with a reference, this gives a curve showing reflected energy. By aligning the minima of the curve with the target frequency and ensuring that the minima is as close to zero intensity as possible, the spectrometer can easily be tuned.
Watch this video for a demonstration of tuning and matching in the X-Pulse: Measuring 7 nuclei in 37 minutes

Figure 3 Tuning & Matching in Oxford Instruments SpinFlow software, with the 1H/19F channel in red and the X-channel in blue: (a) both channels well-tuned & matched; (b) 1H/19F channel well-tuned but poorly matched; (c) X-channel poorly tuned but well matched; (d) 1H/19F channel poorly tuned & matched.
Since the electrical properties of samples vary, optimal tuning & matching of an NMR spectrometer varies between samples. These differences arise mostly from the solvent, and the degree of ionic character of the sample. The most significant changes are observed when changing between ionic aqueous solutions and low-polarity organic solvents. Therefore, it is important to tune & match on individual samples, and not just on the individual nuclei.
While the vast majority of elements have at least one NMR active nuclei and in principle can be observed by NMR Spectroscopy, there is a number of considerations related to acquiring spectra in practice. This includes the NMR-active nuclei's abundance and sensitivity, and if it is possible to tune the NMR spectrometer to the appropriate frequency. To obtain spectra of different nuclei it's important to be able to tune the NMR spectrometer not only to the individual nuclei, but for the specific sample.
The Oxford Instruments X-Pulse 90/60 Broadband Benchtop NMR Spectrometers are fully tuneable systems, capable of observing a wide range on nuclei including 1H, 19F, 13C, 31P, 7Li, 11B, 23Na, 27Al, 29Si and many more.