Segregation—the uneven distribution of alloying elements or impurities in steel—is an unavoidable thermodynamic process that occurs during casting. As the alloy cools, element migration can occur, and without correction these segregations become locked into the final product1. If significant, there could be local changes to the mechanical and corrosion response2,3,4.
Detecting and understanding segregation is therefore essential to guarantee product performance and ensuring manufacturing consistency. While light microscopy may reveal segregation in some cases, heat treatments can obscure or remove visible indicators. Direct chemical analysis using wavelength-dispersive X-ray spectroscopy (WDS) on an electron microprobe can identify and quantify minor (< 0.5 wt.%) segregations. Though commonplace, this technique is time consuming, requires specialist operation, and comes at significantly increased cost in comparison with SEM-based techniques.
Here, the Unity BEX (Backscattered Electron and X-ray) detector offers a new solution. Its high sensitivity and rapid mapping capability make it possible to detect minor chemical variations across large areas in a fraction of the time required by traditional methods. Working seamlessly within a scanning electron microscope (SEM), BEX delivers both the speed and precision required for practical segregation detection—while fully correlated EDS provides quantitative validation. This combination offers manufacturers and researchers an accessible, cost-effective, and powerful approach to understanding segregation along the entire process chain.
Two commercially available austenitic stainless steels (SS316L and SS304L) were analysed using the Unity BEX detector on a high-resolution field-emission SEM. BEX maps were acquired at 20 kV and 3 nA, with six individual 300× magnification maps collected per sample in just 20 minutes.

Figure 1 – Large-area BEX montages of stainless steels SS316L and SS304L reveal striking differences in segregation patterns. SS304L, manufactured as a plate, exhibits significantly more pronounced segregation than SS316L, manufactured as a bar—highlighting how processing history directly influences microstructural uniformity.
The results suggest that manufacturing history can produce different segregation behaviour. This rapid visualisation provided an immediate, qualitative insight into compositional differences that could impact steel performance.
A welded steel sample was analysed over a 4 cm² area to determine micro-segregation patterns, namely Ni-Cr variations, and the influence of welding heat. Using 20 kV and 3.5 nA, 239 mapped fields were automatically collected in 13 hours. The resulting montage contained 734 million pixels, producing a high-resolution dataset for in-depth analysis. Simultaneously, a fully correlated EDS dataset was acquired using a single Ultim Max ¥ 100.

Figure 2 – Large-area (4 cm²) BEX mapping of a welded SS316L sample shows the clear boundary between heavily segregated base steel and homogeneous weld metal. The undulating fusion line marks where segregation ceases, providing immediate visual confirmation of microstructural change across the heat-affected zone (HAZ).
The BEX image revealed a well-defined fusion line separating heavily segregated steel from the more homogeneous weld material. Segregation patterns extended over centimetres in the steel, while high-magnification extracts at 1 mm and 8 cm from the fusion line confirmed that these patterns pre-dated the welding process, occurring both within and outside of the heat affected zone.
Using the reconstruct spectrum tool, three EDS spectra were extracted from the correlated EDS map within the data cube. Guided by the qualitative BEX element map (Fig. 3A), EDS spectrum reconstruction revealed up to 5 wt.% variation in Cr and Ni within the steel (Fig. 3B). This variation became immediately apparent after QuantMap processed the BEX data, generating a fully quantified 2D representation of Ni distribution across the sample (Fig. 3C).

Figure 3 – (A & B) Quantitative EDS extracted from the large-area BEX dataset shows substantial variation in Cr and Ni content within the segregated steel—up to 5 wt.%. (C) QuantMap analysis of a 2 mm² subsection displaying Ni wt.%. The map reveals more than 6 wt.% variation in Ni concentration as a result of segregation.
This variation has practical consequences: localised differences in Ni content could have triggered the formation of residual ferrite and alter the steel’s corrosion/mechanical behaviour, potentially leading to uneven corrosion resistance. Interestingly, in this case, the Cr content remained above the critical 12% level in the regions studied and no loss in passivation is expected.
Micro-segregation analysis demands both speed and accuracy—requirements that can challenge traditional electron microprobe-WDS workflows. The Unity BEX detector offers a practical solution, delivering rapid qualitative detection with the ability to perform fully correlated quantitative maps. Furthermore, our SEM-BEX approach can be performed at greatly reduced cost suitable for less experienced microanalysis.
In this study, combined BEX and EDS enabled:
By combining fast, high-count BEX mapping with fully correlated EDS analysis, manufacturers and researchers can detect, visualise, and quantify segregation efficiently, helping ensure material performance and reliability from production through to application.