SDK Index Values for Readout Speeds

Author: Andor Support Team

Published: 01 Apr 2022 · Last updated: 13 Apr 2022

Tags: support-resource

This article outlines the structure of how shift speeds correspond to index values required to select them.

Vertical shift speeds will index from fastest, being ‘0’, to slowest using the function SetVSSpeed.

Horizontal readout speeds will index from fastest, being ‘0’, to slowest using the function SetHSSpeed; but will also depend on the AD Channel selected. Different AD channels will have different horizontal shift speeds available (SetADChannel).

For example, iXon3 DU-897E

SetADChannel Index SetHSSpeed Index
16-bit 0 1MHz 0
14-bit 1 3MHz 2
14-bit 1 5MHz 1
14-bit 1 10MHz 0
SetVSSpeed Index
0.3us 0
0.5us 1
0.9us 2
1.7us 3
3.3us 4

To return the number of vertical shift speeds: GetNumberVSSpeed.
To return the number of horizontal shift speeds: GetNumberHSSpeed.
To return the number of AD channels: GetNumberADChannel.