Measuring the oil content of dried snack foods is essential for quality control to ensure that the products meet their nutritional values specifications. The amount of oil used is also important in terms of the significant cost of the raw material and the effect it may have on the texture and perceived quality of the product.
Solvent extraction techniques are commonly used for determination of fat content. However, they tend to be slow, laborious, inaccurate and require skilled personnel. In addition, many of the often hazardous chemicals used are becoming increasingly unacceptable according to international environmental standards. Despite these issues solvent extraction continues to be used as a reference measurement for quality control.
Instrumental methods are often referred to as secondary techniques since they are usually set up to match the results produced by solvent extraction. To provide a result equivalent to the traditional extraction techniques, secondary techniques require a calibration against the reference technique used. Although they are fast and easy to maintain, many secondary techniques need to be calibrated and maintained regularly. Maintenance and consumables also add significantly to the cost of ownership. For example, although Supercritical Fluid Extraction (SFE) is reasonably fast, it requires high maintenance and the cost of compressed CO2 used to extract oil is also significant. Near Infra-Red (NIR) is commonly used for on-line monitoring but is difficult to apply on opaque samples as it can only scan the surface. It is also complex to calibrate as measurements are sensitive to product granularity and spices, and therefore it is difficult to maintain accurate results on a large variety of product types. For that primary reason, NIR has limited applicability for the quality control of oil content in snack foods.
In contrast, Nuclear Magnetic Resonance (NMR) provides a simple, accurate and inexpensive measurement of oil content in snack foods which is routinely used in factories around the world.
Benchtop NMR detects the signal from oil and distinguishes it from signals from solids and bound water (<10% moisture).
Although a primary calibration can be produced using a single sample of oil (100%), Calibration Maintenance Samples (CMS) that are stable long term can be allocated to reference concentrations (nominally 10, 25 and 40%). Subsequently the mid-range SUS is used for routine checks to ensure that an instrument provides results within acceptance limits. Should the result fall outside these limits, the low and high SUSs are used to recalibrate the instrument.
The results are comparable to those of acid hydrolysis followed by soxhlet extraction (e.g. Weibull-Stoldt) which provides a measurement of the total amount of oil. Should the user require the results to match those from Soxhlet, they will need to measure a representative set of samples using both NMR and Soxhlet to determine the skew and bias from the 100% oil calibration thus in the long term only one calibration needs to be maintained.
A sample vial is tared then filled to a given height with the sample using a dedicated packing tool. Each sample is then weighed before being placed in a conditioning block for 20 minutes. The conditioning temperature is usually 40°C, the same as the magnet, for optimum precision. For some oils, particularly palm oil, the conditioning temperature needs to be higher to ensure the oil has totally melted.

Fig. 1 — Correlation between NMR signal/mass and reference values for oil content in snack foods
Table 1 summarises the results of several samples that were selected according to their characteristic compositions and manufacturing process. The oil contents determined by NMR are compared against those determined by the reference technique, in this case solvent extraction. The NMR data (signal/mass) plotted against those reference values in Figure 1 demonstrates the accuracy of the NMR technique as all the data points align on a straight line and the results are relative to the NMR signal of 100% corn oil from which the samples were produced.
| Snack Product | Ref % | NMR (%) | Diff (%) |
| A crushed | 0.83 | 0.91 | 0.08 |
| A ground | 0.83 | 1.00 | 0.17 |
| B crushed | 5.59 | 5.46 | 0.13 |
| B ground | 5.59 | 5.64 | 0.05 |
| C crushed | 11.07 | 11.00 | -0.07 |
| C ground | 11.07 | 11.04 | -0.03 |
| E ground | 22.18 | 22.44 | 0.26 |
| F ground | 22.98 | 23.29 | 0.31 |
| G crushed | 25.58 | 25.02 | -0.56 |
| G ground | 25.58 | 25.11 | -0.47 |
| H crushed | 26.51 | 26.60 | 0.09 |
| H ground | 26.51 | 26.60 | 0.09 |
| I crushed | 33.96 | 34.06 | 0.10 |
| I ground | 33.96 | 33.68 | -0.28 |
| J ground | 34.84 | 35.01 | 0.17 |
| H crushed | 43.90 | 43.66 | -0.24 |
| H crushed | 43.90 | 44.34 | 0.44 |
| Corn oil product line 1 | 100.00 | 99.80 | -0.20 |
| Corn oil product line 2 | 100.00 | 100.37 | 0.37 |
| Standard error of estimate (%) | 0.23 |
Table 1: Comparison between NMR results and reference values obtained from solvent extraction
Table 1 also shows that there is no significant effect whether the samples have been crushed or thoroughly ground. Also, when different oil types are compared, there is no significant difference between fresh or processed corn oil. Table 2 shows that the repeatability (or precision) of the measurement on the same sample is very good. Table 3 shows there is only a small difference in the apparent oil content between corn, sunflower, cottonseed and even palm oil owing to differences in their hydrogen density.
| Measurement Number | Oil content by NMR (%) |
| 1 | 34.74 |
| 2 | 34.69 |
| 3 | 34.80 |
| 4 | 34.70 |
| 5 | 34.77 |
| 6 | 34.69 |
| 7 | 34.72 |
| 8 | 34.79 |
| 9 | 34.74 |
| 10 | 34.71 |
| Standard deviation (%) | 0.04 |
| Precision (2s, %) | 0.09 |
Table 2: Repeatability measurements by NMR
| Oil type | Oil content by NMR (%) |
| Corn | 100.00 |
| Cotton seed | 99.12 |
| Sunflower | 99.59 |
| Palm | 99.95 |
* For optimal precision samples should be conditioned at 40°C for 20 minutes in a dry block heater prior to analysis.
The MQC+23 with a 0.55 Tesla (23 MHz) magnet, fitted with a 26 mm diameter (10 ml sample) probe is ideal for this application. The Oil in Snack Food package comprises:
In addition you may require:
+ Not including sample conditioning time