The study focuses on the chemical characterisation of particulate material captured by an in-line liquid chromatography filter, where the recessed filter position and surrounding housing obstruct the X-ray path to a standard inclined EDS detector (Figs. 1&2). In these situations, a backscattered electron and X-ray (BEX) detector provides an alternative analytical approach, making in-situ analysis possible without destructive disassembly. This is particularly valuable for filter inspection, where identifying the composition and morphology of retained particles can help diagnose contamination pathways, protect instrument performance, and support quality control in highly regulated industries such as pharmaceuticals.

Figure 1: Photographs displaying the filter and housing analysed by BEX. Note from the top view that the filter disc is significantly lower than the surrounding housing, which leads to X-ray shadowing when using a conventional EDS detector.
Specifically, a BEX detector offers a unique solution to this problem because of its horizontal insertion beneath the SEM pole piece (Fig. 2). As a result, the X-ray line of sight is not obstructed by the filter housing, unlike with EDS detectors that are typically inclined at 35° (Fig. 2). This means that high-quality in-situ chemical characterisation is possible within the SEM for many samples with challenging geometry, without the need to remove the filter housing, preserving the samples integrity.
Figure 2: Diagrams displaying the shadowing effect caused by using an inclined conventional EDS detector by comparison with a BEX detector inserted horizontally beneath the pole piece.
Operated at beam conditions of 20 kV and 5 nA, the BEX detector was able to identify Fe- and Si-rich particulates captured within an in-line liquid chromatography filter after just 40 seconds of acquisition time (Fig. 3). Particulates as small as 1.7 µm were easily identified from a single high-resolution (8K) map (Fig. 3C). The composition of these particles suggests that they originate from the physical breakdown of Si-rich glassware and mechanical wear of the system’s metallic hardware.

Figure 3: BEX analysis of particulate contamination retained within an in-line filter. A) Low magnification 8K BEX map of the filter area, approximately 1 mm wide, captured in 40 seconds. B) Fe X-ray map acquired using the BEX detector, extracted from the multilayer elemental and BSE dataset shown in A. C) Higher-magnification view from map A showing Si-rich particulate contamination. D) Higher-magnification view from map A showing Fe-rich particulate contamination, including particles several µm in size.
This case study highlights how samples with challenging geometry can prevent effective chemical characterisation by SEM-EDS because of shadowing. By overcoming this limitation through horizontal insertion, the BEX technique enables rapid compositional analysis in situations where conventional EDS struggles. As a result, the origin of particulate contamination within a high-performance liquid chromatography system can be identified without destructive sample preparation and risk of sample contamination.